











|
 |
|
|
CIVM research
|
|
|
Wilfried Mai (former visiting researcher from Bernard University, Lyon, France ), CT Badea, CT Wheeler, LW Hedlund, G. Allan Johnson (Center for In Vivo Microscopy)
|
Effects of breathing on spatial resolution in microscopic imaging of rodents
|
Information extracted from poster presented at 2004 IEEE International Symposium on Biomedical Imaging: From Nano to Macro, Arlington, VA, April 15-18, 2004. See the pdf of the
article about this project: W Maï, CT Badea, CT Wheeler, LW Hedlund, GA Johnson, Effects of breathing and cardiac motion on the spatial resolution in microscopic imaging of rodents,
Magn Reson Med 53(4): 858-865, 2005.
| Introduction
MR microscopy can produce high-resolution pulmonary images in live rodents by synchronizing the image acquisition across multiple breaths [1]. The precision to control motion will most likely define the attainable resolution limits. This work evaluates how reliably the respiratory structures return to the same position from breath to-breath each time data are acquired.
|
| Methods
Radio-opaque tungsten beads (0.28-mm diameter) were surgically glued on the abdominal surface of the diaphragm (Figure 1) of 10 anesthetized rats that were mechanically ventilated using a custom-built ventilator [2].
The rats were mechanically ventilated at 60 breaths/min (tidal volume: 1.83.0 ml). A solid-state pressure transducer on the breathing valve measured airway pressure.
Pediatric electrodes were taped on the footpads for ECG. These signals were also used to control the biological pulse sequences described below. The range of motion of the beads (relative to a reference vertebral bead) was evaluated using digital micro-radiography, by measuring the standard deviation of the distance between these beads and the reference bead on a series of 150200 consecutive images.
We used two specific biological pulse sequences (Figures 2a and 2b)ventilation-synchronous acquisition, and both ventilation-synchronous and cardiac-gated acquisition. The control and acquisition system was made up of three computers, a high-capacity angiography tube with 0.1-mm focal spot driven by a Phillips CXP 80 generator, and a 50-micron digital camera (Microphotonics XQUIS, Photonics Science, East Sussex, UK) (Figure 3). The programs that controlled the x-ray tube and camera according to the physiological parameters were developed using LabVIEW (National Instruments, Austin, TX).
|
Figure 1: Ventro-dorsal projection of the thorax in a rat. Note the three implanted beads on the diaphragm and a reference bead on a thoracic vertebra. The insert shows a close-up view of the left bead.
|
Figure 2a: Ventilation synchronous acquisition. For breath-hold images, the x-ray pulse occurred 150 (early breath-hold) to 250 ms (late breath-hold) after the beginning of inspiration (shown by dark-gray). For end-expiratory images, the x-ray pulse occurred 850 ms after the beginning of the inspiration. The camera was enabled at the same time as the x-ray pulse.
|
Figure 2b: Ventilation-synchronous and cardiac-gated images. A 110-ms acquisition window was defined (here represented for the breath-hold images) and any QRS occurring within this window triggered an x-ray pulse and enabled the data acquisition by the camera.
|
|
Figure 3: Schematic of the acquisition system controlled by three computers. One computer controls the ventilator and acquisition of physiological parameters. Another computer controls the x-ray rotor, enables the camera, and generates the x-ray pulse according to the defined biological pulse sequence. The third computer records the digital data coming from the camera.
Results
Table 1 lists the mean difference in the distances between inspiratory breath-hold and end-expiration, as measured on ventro-dorsal and cross-table projection. Due to proximity to the heart, the bead on the dome of the diaphragm was believed to demonstrate a different behavior regarding its motion over time.
|
Table 1: Mean differences (in microns) in distances between inspiratory breath-hold and end-expiration images, which represents the diaphragm excursion over the breathing cycle, providing an insight into the actual spatial resolution that would be achieved without control of breathing motion.
Table 2: Global results. For each biological pulse sequence, we present the standard deviation in microns averaged across all animals. Values listed are averages of the values obtained on the two perpendicular projections.
|
The results for this bead ( dome) are therefore shown individually compared to the values obtained for the beads glued on the dorso-lateral parts of the diaphragm ( lateral), which were averaged. For each animal, we obtained a standard deviation of the dome and lateral beads motion for a sequence of 150200 images.
Table 2 reports these results averaged across the multiple experiments for 10 animals.
Conclusion
This work provides the first quantitative measure of motion-imposed resolution limits for in vivo imaging. The standard deviation of the displacement was ≤100 microns, which is comparable to the resolution limit for in vivo magnetic resonance imaging imposed by signal-to-noise constraints [1]. As methods for improving the sensitivity develop, methods to limit the impact of motion must follow. As demonstrated here, with careful control of motion, its impact on resolution can be limited.
| References |
|
- GA Johnson, GP Cofer, LW Hedlund et al., Magn Reson Med, 45: 365-370, 2001.
|
2. LW Hedlund, GA Johnson, ILAR J 43: 159-174, 2002.
|
Acknowledgements All work was performed at the Duke Center for In Vivo Microscopy, an NCRR National Resource (P41 RR05959). Additional support provided by NCI (R24 CA-92656) and the (DoD-556450). The author supported in part by the French Ministry of Foreign Affairs (Lavoisier grant).
|
|
|